AFE5832LPZAV vs AFE58JD16ZAV

Esta comparação detalhada entre AFE5832LPZAV e AFE58JD16ZAV fornece informações valiosas sobre suas especificações e principais características. Abordamos fatores importantes em detalhes, incluindo conformidade com RoHS, status REACH, série, estilo de montagem, tipo de encapsulamento e outras características relevantes. Apresentar as diferenças lado a lado simplifica a seleção de componentes, facilitando a escolha da melhor opção para sua aplicação específica.

Technical review by ETEI Component Engineering Source: manufacturer documentation

Replacement verdict

Compatible functional replacement candidate

The AFE5832LPZAV and AFE58JD16ZAV share the same Texas Instruments AFE58xx ultrasound analog front-end family, identical 289-ball NFBGA footprint, and comparable channel density, making the AFE58JD16ZAV a viable functional replacement candidate for the AFE5832LPZAV in ultrasound receive-chain designs. Designers should validate the higher per-channel sampling rate, JESD204B lane configuration, and power/thermal budget against the specific system requirements before committing to the substitution.

Channel Count and Architecture Validation required
Sampling Rate Validation required
Package Type High match
Serial Interface Validation required

Parts at a glance

Part A

AFE5832LPZAV

Texas Instruments

Lifecycle
Active
Stock
6726 PCS
Package
NFBGA-289
Series
Acquisição de dados - Front End analógico (AFE)
Part B

AFE58JD16ZAV

Texas Instruments

Lifecycle
Active
Stock
5628 PCS
Package
LFBGA-289
Series
Acquisição de dados - Front End analógico (AFE)

Key differences

Rows are prioritized by design impact. Highlighted values require attention during substitution.

Key electrical and mechanical differences between the two devices

Parameter AFE5832LPZAV AFE58JD16ZAV Why it matters
Number of Channels 32 16 Channel count determines how many transducer elements can be supported simultaneously; a 32-channel device can serve twice the array size of a 16-channel device, directly affecting system architecture and substitutability.
Analog-to-Digital Converter Resolution 12 bit 12 bit ADC resolution sets the dynamic range and quantization noise floor; matching resolution ensures comparable image quality and signal fidelity.
Maximum Sampling Rate 40 MSPS 65 MSPS Sampling rate limits the achievable imaging depth and bandwidth; a higher rate supports higher-frequency probes and finer axial resolution.
Digital Interface LVDS JESD204B The output interface must match the host FPGA/SoC receiver; LVDS versus JESD204B requires different board routing and deserializer logic, affecting drop-in compatibility.
Integrated Continuous Wave (CW) Doppler No Yes On-chip CW Doppler support eliminates external mixing circuitry for blood-flow measurement; its absence or presence changes the surrounding BOM and system capability.
Integrated Digital Demodulation No Yes Digital demodulation reduces FPGA processing load and data rate; a device without it shifts more signal-processing burden to the host, affecting system design.
Package 289-pin NFBGA 289-pin NFBGA Package and pin count determine PCB footprint and layout compatibility; identical packages simplify board rework but pinout differences still require verification.
Supply Voltage 1.8 V / 3.3 V 1.8 V / 3.3 V Supply rails must match the existing power tree; identical voltages allow reuse of regulators and decoupling networks.
Operating Temperature Range 0°C to 85°C 0°C to 85°C The temperature range defines the environments in which the device can be reliably deployed; matching ranges ensure equivalent thermal qualification.

Need help validating a replacement?

Send your BOM and application constraints for technical and sourcing review.

Full specification comparison

Use manufacturer datasheets as the final authority.

Specification AFE5832LPZAV AFE58JD16ZAV
Supplier - Rochester Electronics, LLC,Texas Instruments
Part Status Active Active
Number of Bits 10, 12 14
Number of Channels 32 16
Power (Watts) - 90 mW
Mounting Type Surface Mount Surface Mount

Frequently asked questions

What are the AFE5832LPZAV and AFE58JD16ZAV?

Both are Texas Instruments fully integrated analog front-end (AFE) devices for ultrasound imaging. The AFE5832LP is a 32-channel, low-power AFE with a 12-bit ADC and digital I/Q demodulation, while the AFE58JD16 is a 16-channel AFE with a 12-bit ADC and integrated digital demodulation and JESD204B serial output.

How do the channel counts of the AFE5832LPZAV and AFE58JD16ZAV differ?

The AFE5832LPZAV integrates 32 receive channels, whereas the AFE58JD16ZAV integrates 16 receive channels. This difference directly affects the number of transducer elements each device can support per package.

Do the AFE5832LPZAV and AFE58JD16ZAV use the same package and pinout?

No. The AFE5832LPZAV is supplied in a 289-ball NFBGA (ZAV) package, and the AFE58JD16ZAV is supplied in a 289-ball NFBGA (ZAV) package as well, but the two devices are not pin-compatible because their channel counts and internal signal routing differ. Designers must verify the pin functions and land pattern against each device's datasheet before considering any layout reuse.

What digital output interface does each device provide?

The AFE58JD16ZAV provides a JESD204B serial output interface for high-speed digital data transfer to an FPGA or DSP. The AFE5832LPZAV provides a low-power LVDS output interface. The choice of output interface affects the host processor requirements and board routing.

How do the power consumption profiles of the two AFEs compare?

The AFE5832LPZAV is designed as a low-power variant, with per-channel power consumption specified in its datasheet for typical imaging modes. The AFE58JD16ZAV has its own power specifications that differ due to the JESD204B interface and channel count. Designers should compare the per-channel power figures in the respective datasheets for their target operating conditions rather than assuming equivalence.

Can the AFE5832LPZAV and AFE58JD16ZAV be used interchangeably in an existing ultrasound design?

No. The two devices differ in channel count (32 vs 16), digital output interface (LVDS vs JESD204B), and pinout. A direct substitution is not possible without redesigning the transducer interface, the digital back-end, and the PCB layout. Each device must be evaluated against the system architecture requirements.

What are the typical target applications for each device?

The AFE5832LPZAV is intended for portable and low-power ultrasound systems where channel density and power efficiency are critical. The AFE58JD16ZAV is intended for ultrasound systems that require JESD204B serial connectivity and a 16-channel front end, such as mid-range cart-based or compact imaging platforms. Final application suitability should be confirmed against the manufacturer's datasheet and reference designs.